To use this function you have to enable JavaScript and allow popup windows for this domain.
 
Analytical Columns

 

Pro Family : YMC-Pack Pro C18 RS
 

The stability properties make YMC-Pack Pro C18 RS an excellent choice for the separation of non-polar structurally related analytes. The extended resistance towards acidic and basic conditions not only widens the possibilities in method development but also provides further selectivities for demanding separations such as LC-MS or combinatorial chemistry of: positional isomers, large hydrophobic molecules, basic and acidic compounds, peptides

Separation of peptides under LC-MS compatible conditions
Column: YMC-Pack Pro C18 RS (5 μm, 8 nm, 150 x 4.6 mm)
Eluent: A: water / TFA (100/0.1)
B: acetonitrile / TFA (100/0.1)
Gradient: 15-35% B (0-15 min)
Flow: 1.0 ml/min
Detection: UV, 220 nm, 0.32 AUFS
Temperature: 37 ºC
Injection: 20 μl (0.05 mg/ml)
Separation of acidic and basic drugs under
LC-MS compatible conditions
Column: YMC-Pack Pro C18 RS (5 μm, 8 nm, 150 x 4.6 mm)
Eluent: 20 mM CH3COONH4-NH3 (pH 8.9) / methanol (35/65)
Flow: 1.0 ml/min
Detection: UV, 220 nm, 0.64 AUFS
Temperature: 30 ºC
Injection: 10 μl (0.1 mg/ml, 0.036 μl/ml)
Acidic compounds
Food preservatives
Column: YMC-Pack Pro C18 RS (150 x 4.6 mm)
Eluent: 20 mM CH3COONa - CH3COOH
(pH 4.3) / acetonitrile (80/20)
Flow: 1.0 ml/min
Detection: UV at 230 nm, 0.26 AUFS
Temperature: 37 ºC
Injection: 10 μl (0.02 mg/ml)
Food preservatives
Chlorophenol isomers
Column: YMC-Pack Pro C18 RS (150 x 4.6 mm)
Eluent: acetonitrile / water / acetic acid
(30/70/1)
Flow: 1.0 ml/min
Detection: UV at 230 nm, 0.32 AUFS
Temperature: 37 ºC
Injection: 10 μl (0.1 - 0.2 mg/ml)
Nicardipine hydrochloride
Column: YMC-Pack Pro C18 RS (150 x 4.6 mm)
Eluent: 20 mM KH2PO4-K2HPO4 (pH 6.9) /
methanol (25/75)
Flow: 1.0 ml/min
Detection: UV at 254 nm, 0.32 AUFS
Temperature: 37 ºC
Injection: 10 μl

For more applications please refer to our “Application Data Collections” or contact us directly.