The stability properties make YMC-Pack Pro C18 RS an excellent choice for the separation of non-polar structurally related
analytes. The extended resistance towards acidic and basic conditions not only widens the possibilities in method development
but also provides further selectivities for demanding separations such as LC-MS or combinatorial chemistry of:
positional isomers, large hydrophobic molecules, basic and acidic compounds, peptides
| Separation of peptides under LC-MS
compatible conditions |
 |
Column: YMC-Pack Pro C18 RS (5 μm, 8 nm, 150 x 4.6 mm)
Eluent: A: water / TFA (100/0.1)
B: acetonitrile / TFA (100/0.1)
Gradient: 15-35% B (0-15 min)
Flow: 1.0 ml/min
Detection: UV, 220 nm, 0.32 AUFS
Temperature: 37 ºC
Injection: 20 μl (0.05 mg/ml) |
Separation of acidic and basic drugs under
LC-MS compatible conditions |
 |
Column: YMC-Pack Pro C18 RS (5 μm, 8 nm, 150 x 4.6 mm)
Eluent: 20 mM CH3COONH4-NH3 (pH 8.9) / methanol (35/65)
Flow: 1.0 ml/min
Detection: UV, 220 nm, 0.64 AUFS
Temperature: 30 ºC
Injection: 10 μl (0.1 mg/ml, 0.036 μl/ml) |
| Acidic compounds |
 |
Food preservatives
Column: YMC-Pack Pro C18 RS (150 x 4.6 mm)
Eluent: 20 mM CH3COONa - CH3COOH
(pH 4.3) / acetonitrile (80/20)
Flow: 1.0 ml/min
Detection: UV at 230 nm, 0.26 AUFS
Temperature: 37 ºC
Injection: 10 μl (0.02 mg/ml) |
| Food preservatives |
 |
Chlorophenol isomers
Column: YMC-Pack Pro C18 RS (150 x 4.6 mm)
Eluent: acetonitrile / water / acetic acid
(30/70/1)
Flow: 1.0 ml/min
Detection: UV at 230 nm, 0.32 AUFS
Temperature: 37 ºC
Injection: 10 μl (0.1 - 0.2 mg/ml) |
| Nicardipine hydrochloride |
 |
Column: YMC-Pack Pro C18 RS (150 x 4.6 mm)
Eluent: 20 mM KH2PO4-K2HPO4 (pH 6.9) /
methanol (25/75)
Flow: 1.0 ml/min
Detection: UV at 254 nm, 0.32 AUFS
Temperature: 37 ºC
Injection: 10 μl |
For more applications please refer to our “Application Data Collections” or contact us directly.
|