Analytical Columns
Pro Family : YMC-Pack Pro C4
 

General
More than 80% of the RP-HPLC analyses are accomplished on octyl- or octadecyl-phases. Because of this overwhelming majority, many chromatographers neglect other selectivities that might be better suited to their separation, such as butyl phases. With Pro C4, YMC offers a stationary phase based on the well-known ultra pure silica of the ProFamily. Compared to a C18-phase with the same eluent, this less hydrophobic material gives shorter retention times for non-polar compounds while the retention time of polar analytes are virtually unaffected. This makes the Pro C4 an interesting alternative especially when short analysis times are required.
For this reason, mixtures with a wide range of component polarity are best analysed by short chains, such as YMC-Pack Pro C4.
The comparison shown on the right demonstrates that YMC-Pack Pro C4 is the column of choice when fast HPLC is required. There is almost no difference in retention times for the first three compounds whilst Ibuprofen elutes faster on YMC-Pack Pro C4 due to its lower polarity.

 
Separation of pain-killing drugs on YMC-Pack
Pro C18, Pro C8 and Pro C4
Column: 150 x 4.6 mm i.d., 5 µm, 12 nm
Eluent: methanol / 20mM H3PO4 (60/40)
Flow rate: 1.0 ml/min
Temperature: 37° C
Detection: UV at 254 nm, 0.32 AUFS
Injection: 10 µl (0.01 - 2.0 mg/ml)

 

Column care
YMC-Pack Pro C4 is stable towards hydrolysis between pH 2-7,5. Remove acid and buffer salts before storage. Store the column in methanol / water = 70/30. Clogged inlet frits often can be cleaned by changing the flow direction or replacement.
For detailed information please refer to the “Column Care and Use Instructions”, which are shipped with each analytical column.


 